Endurance
10th February 2022 Endurance Main 2.02.31 (includes changes and improvements from pre-release candidate 2.02.30) New ASCII Commands and parameters s...
Thu, 19 Jan, 2023 at 1:38 PM
Once you replace the old Fiber Cable with the newly ordered Fiber Cable of the Endurance sensor, Fiber Calibration needs to be performed. Each replacem...
Thu, 1 Sep, 2022 at 6:51 AM
Changing the Window Sometimes extremely harsh environments can cause damage to the window. A replacement window (E-PW) is available from the manufacturer....
Wed, 27 Jul, 2022 at 5:52 PM
Dimensions or information not found in manual : Sapphire window dimensions E-BF1 & E-FORQP:
Tue, 2 Feb, 2021 at 11:23 AM
If you experience a reset to factory default settings in the Endurance during a power cycle and can isolate that it is not caused by Ethernet/IP or Profinet...
Wed, 2 Mar, 2022 at 4:32 PM
A very small number of users globally have reported the sensor reverted to degrees C from degrees F. A repeatable pattern across the user base has not ...
Fri, 21 Jan, 2022 at 4:45 PM
Endurance differs from all other devices we have, in that the focus distance does not start from the front face of the device. The front of the enduranc...
Tue, 23 Nov, 2021 at 4:18 PM
Temperature Limits for related Error codes ECHH >= 80°C ECUU <= 50°C EIUU <= 0°C EIHH >= 85°C
Fri, 15 Oct, 2021 at 3:53 PM
This page consolidates extracts from the Modline 5 and Endurance manuals. The E-5M5PK is approximately 30cm in length. There is a discrepancy betwee...
Tue, 6 Apr, 2021 at 10:58 AM
Download the 8 page Endurance accessories and spare parts (sales) document Endurance Accessories
Tue, 6 Apr, 2021 at 10:41 AM