Once you replace the old Fiber Cable with the newly ordered Fiber Cable of the Endurance sensor, Fiber Calibration needs to be performed. 


Each replacement fiber optic cable is calibrated at the factory before shipping. The calibration constants are sent along with a label mounted on the cable. So you have to enter them into the appropriate <Change Fiber> submenu of the DataTemp Multidrop software program. This program sends the new calibration constants to the sensor’s electronics.

The DataTemp Multidrop software comes with the data carrier you received. To run the program and enter new cable calibration constants, complete the following:


1. Install and run the DataTemp Multidrop software.

2. Establish communication between the device and the software.

3. Navigate to the main menu <Devices>, and select the submenu <Change Fiber>.

4. In the following dialog you are requested to input the calibration constants for the fiber cable.