Base Model
Description and Specs
Remarks
T40-CO2-30-SF0-X
Photodetector type             : Detector TS80 Thermopile
Spectral Response : 4.24 µm
Operation temp range : Flame Detection 120 to 1650°C
Optical resolution D:S : 33:1
Response time : 130ms
Repeatability : ± 0.5% of reading for Tmeas > 700°C
± 1% of reading or ± 3°C for Tmeas ≤ 700°C
System Accuracy : ± 1.0% of reading for temperature > 700°C
± 2.0 % of reading or ± 6 K for Tmeas ≤ 700°C
 Focal Distance                   : SF0 1520mm (60 in)
Measurement of CO2 Gas Flames
Suitable to replace
D54TX....
DS6TX...
D95TX... (ABB special)
D49TX..
DR1TX....(check that 1650°C is acceptable as this special had 1750°C top temp'

T40-CO-30-SF0-X

Photodetector type                 : Detector TS80 Thermopile

Spectral Response                 : 4.64 µm

Operation temperature range : Flame Detection 120 to 1650°C

Optical resolution D:S             : 33:1

Response time                        : 130ms 

Repeatability                           : ± 0.5% of reading or ± 2.5°C 

System Accuracy                    : ± 1.0% of reading or ± 5 K 

Focal Distance                        : SF0 1520mm (60 in)

Measurement of CO Gas Flames
Suitable to replace 
D52TX....
T40-NOX-30-SF0-X

Photodetector type                  : Detector TS80 Thermopile

Spectral Response                  : 4.47 µm

Operation temperature range  : Flame Detection 120 to 1650°C

Optical resolution D:S              : 33:1

Response time                         : 130ms 

Repeatability                            : ± 0.5% of reading 

System Accuracy                     : ± 1.0% of reading + 2°C 

Focal Distance                         : SF0 1520mm (60 in)

Measurement of NOx Gas Flames
Suitable to replace 
D51TX....
DRYTX....


Special note: When replacing D51, D52 and D54 TX sensors with these T4.0 version it will necessary to re-scale the anlog output to start at 250°C). USB Micro - USB A cable and DTMD software needed


Additional Useful Technical Info Links:


T4.0 How To Scale Analog Output


Thermalert T4.0 How to Configure Analog Output Mode


Additional Useful Sales Info and Link:

The like for like replacement model ends with a -0 (2 wire sensor)

However, all of these Flame or Flue Gas versions are available with other output types (4-wire, mA, mV, t/c, Profinet, Ethernet/IP)


Thermalert T4.0 - Considerations when replacing TX and XR