Note: above GIF is converted from screen video and is approximate to the real update rate
1. Only applicable to 4 wire Profinet version.
2. Must use the latest sensor configuration software DTMD v6.1.99
3. It could happen, that the firmware update will not be successful, in which case the previous firmware is not affected. If this happens please try a second time.
Pre-release tested version for solving:
1. Cyclic time - was too long
2. Communication dropouts - related to too long cyclic time
3. "With firmware version 01.03.0319P and the Proneta Siemens software, it is possible to change the device name independently of the IP address (+repower). The IP address change to 0.0.0.0 is applied (+repower).
With firmware version 01.02.1008P, the device name could only be changed if the IP address was set to 0.0.0.0 in Proneta."
4W_Stingray_Profinet_main_01.03.0319.fw attached in this article