Information from FPI Berlin Service Manager


General Overhaul

Check motor for excessive audible bearing noise

Check running time - if close to 45,000 hrs change motor unit

Change IR window, check and adjust

Check onboard I/O (trigger, alarm output, analog output, analog input)


Calibration (certification)

Verification of trigger input and analog output - 3 mA output channels calibrated and adjusted as necessary

 (measured mA values are not included on a calibration certificate per point)

Check IR calibration at 3 points appropriate to the model's temperature range

Advise cost for ISO or DAkkS certification if required.