The attached guide and program will effectively disable the EAAA function in the Ratio sensor.
The EAAA will be shown as EUUU.
Special Note:
The guide also covers modification of the Z command for attenuation failsafe to 99%.
This must be used with caution is it allows the sensor to calculate a temperature in non-deal conditions.
The general recommendation is that this should not be used when peak hold is used (otherwise "high attenuation temperature spikes will be captured based on the leading and trailing edges of a discrete object moving in and out of the FOV).
For DTMD 6.1.99, please send a request to TechSupport Team at the email address: techsupport@flukeprocessinstruments.com
Or Open a ticket at Open Ticket