There are three calibration options for a Datapaq logger. The choice of which calibration to use is based on the requirements you have for your application.


Standard Traceable – This is our standard, traceable calibration performed in our Everett, WA lab. Calibration is valid for one year. The lab is not an accredited lab but the calibration is performed with fully traceable equipment.

 

ISO 17025 Accredited – Same as Standard calibration except it is performed in our Europe ISO 17025 accredited lab. Some companies require calibrations be performed in an ISO 17025 accredited laboratory. This calibration requires a 4-6 week turnaround.

 

AMS2750 (aerospace) /CQI-9 (Automotive) Calibration – These calibrations are one and the same and adhere to AMS2750 and CQI-9 calibration procedures. They would only be selected if required by the customer. Per AMS2750 and CQI-9 requirements, these calibrations are only valid for 3 months.