Please use this table when calibration or checking the most common Endurance sensors
Please note, the column Distance must be used with the following conditions:
For cavity or tube IR calibrators, the distance is to the back of the cavity or tube.
For plate IR calibrators the distance is to an aperture just in front of the plate.
Model | Focus | D:S | Distance | Set focus distance | Aperture |
E1RL | F1 | 100:1 | 625 | 600 | 38 |
E1RL | F2 | 100:1 | 625 | 600 | 38 |
E1RH | F1 | 160:1 | 625 | 600 | 38 |
E1RH | F2 | 160:1 | 625 | 600 | 38 |
E2RL | F1 | 75:1 | 600 | 600 | 38 |
E2RL | F2 | 75:1 | 600 | 600 | 38 |
E1ML | F1 | 160:1 | 600 | 600 | 38 |
E1ML | F2 | 160:1 | 625 | 600 | 38 |
E1MH | F1 | 300:1 | 625 | 600 | 38 |
E1MH | F2 | 300:1 | 625 | 600 | 38 |
E2ML | F1 | 160:1 | 600 | 600 | 38 |
E2ML | F2 | 160:1 | 600 | 600 | 38 |
E2MM | F1 | 160:1 | 600 | 600 | 38 |
E2MM | F2 | 160:1 | 600 | 600 | 38 |
E2MH | F1 | 300:1 | 625 | 600 | 38 |
E2MH | F2 | 300:1 | 625 | 600 | 38 |
E3ML | F1 | 160:1 | 600 | 38 | |
E3ML | F2 | 160:1 | 600 | 600 | 38 |
E3MH | F1 | 300:1 | 600 | 600 | 38 |
E3MH | F2 | 300:1 | 600 | 600 | 8 |
Special Note: Please create a ticket for information on your sensor if it does not appear in this list Create Ticket